Topgram 1:
Lateral, craniocaudalt, API:
inandning
120kV - 35mA
______________________________
Topgram 1:
Lateral, craniocaudalt, API: inandning
120kV - 35mA
_________________________________________________
Serie 1, Calsiumscore seq :
Craniocaudalt, API: inandning
Care kV off 120kV, CareDose4D 80mAs
Slice 3,0, Acq 32 x 1,2, Feed 34.5, Rotation time: Quick 0,28
Scan trigger: ECG, scan.(70 to 70), Pulsing: off, ,Range (70-70)
recon : quick 75ms
___________________________________________
Testdos:
kV:100 mAs:60 coll:1x10,0mm
No scans:30 cycle time:2s, scantime : Full 0,28s
______________________________________________________________
Serie 2: Angio
alt. 1 Flash
Care kV On: Ref.kV 100 (min100kV-max140kV), CareDose4D: Quality ref mAs:360, Slide:11
Acq 128 x 0.6, Pitch: 3.4, Rot time: 0.28, API: Inspiration
Scan trigger: ECG,
alt. 2 Sequence
Care kV on Ref.kV 100 (min100kV-max140kV), CareDose4D: Quality ref mAs:360 , Slide:11
Slice 1,0, Acq 128 x 0.6, Feed 34.0, Rotation time: (0.27s) Flex 0,28s, API:Inspiration
Scan trigger: ECG,
Adaptive Cardio Seq.: on. Scan: 65 to 75, Pulsing: Auto, Range: (70-70) unit% best diastole, Recon: 75ms
själv korrigerar för ev. extraslag och snabbare ryt (breddar nästkommande)
alt. 3 Sequence snabb puls
Care kV on Ref.kV 100 (min100kV-max140kV), CareDose4D: Quality ref mAs:360, Slide:11
Slice 1,0, Acq 128 x 0.6, Feed (34.5), Rotation time: Flex 0,28s, API:Inspiration
Scan trigger: ECG,
Adaptive Cardio Seq.: on, Scan 30-80, Pulsing: Manual, Range:35-72 Unit: %
alt. 4 Spiral
Care kV on Ref.kV 100(min100kV-max140kV), CareDose4D: Quality ref mAs:306, Slide:11
Slice 1,0, Acq 128 x 0.6, Rotation time:(0,28s)
Scan trigger: ECG, API:Inspiration
Pulsing: Auto, Unit:% EST HR.: auto
___________________________________________________________________________________________
_________________________________________________________________________________________________
Coronarartärer XXL:
_________________________
Serie 1, Calsiumscore seq :
Craniocaudalt, API: inandning
Care kV off 120kV, CareDose4D 80mAs
Slice 3,0, Acq 32 x 1,2, Feed 34.5, Rotation time: Quick 0,28
Scan trigger: ECG, scan.(70 to 70), Pulsing: off, ,Range (70-70)
______________________________________________________________________________________________
Serie 2: Angio
alt. 1 Sequence
Care kV on Ref.kV 120 (min100kV-max140kV), CareDose4D: Quality ref mAs:360, Slide:11
Slice 1,0, Acq 128 x 0.6, Feed (34.5), Rotation time: Flex 0,33s
Scan trigger: ECG, API:Inspiration
Scan: 65-75 Pulsing: auto, Unit: %
alt. 2 Sequence snabb puls
Care kV on Ref.kV 120 (min100kV-max140kV), CareDose4D: Quality ref mAs:360, Slide:11
Slice 1,0, Acq 128 x 0.6, Feed (34.5), Rotation time: Flex 0,33s, API:Inspiration
Scan trigger: ECG,
Adaptive Cardio Seq.: on, Scan 30-80, Pulsing: Manual, Range:35-72 Unit: %
alt. 3 Spiral
Care kV on Ref.kV 120 CareDose4D: Quality ref mAs:340, Slide:11
Slice 1,0, Acq 128 x 0.6, Rotation time:(0,28s)
Scan trigger: ECG, API:Inspiration,
Scan range: Auto, Unit:%, EST HR.: auto
_______________________________________________________________________________________________________________________________
_______________________________________________________________________________________________________________
Coronarartärer lågdos unga smala/ 80KV:
________________________________________
Serie 1, Calsiumscore seq :
Craniocaudalt, API: inandning
Care kV off 120kV, CareDose4D 80mAs
Slice 3,0, Acq 32 x 1,2, Feed 34.5, Rotation time: Quick 0,28
Scan trigger: ECG, scan.(70 to 70), Pulsing: off, ,Range (70-70)
___________________________________________________________________________________________________________________
Serie 2: Angio
alt. 1 Flash
Care kV Semi: Ref.kV 80 (min70kV-max120kV), , CareDose4D: Quality ref mAs:560, Slide:11
Acq 128 x 0.6, Pitch: 3.4, Rot time: 0.28, API: Inspiration
Scan trigger: ECG,
alt. 2 Sequence
Care kV Semi Ref.kV 80 (min70kV-max120kV),, CareDose4D: Quality ref mAs:500, Slide:11
Slice 1,0, Acq 128 x 0.6, Feed (34.5), Rotation time: Flex 0,28s?
Scan trigger: ECG, API:Inspiration
Scan: 65 to 75, Pulsing: Auto, unit% ,själv korrigerar för ev. extraslag och snabbare ryt (breddar nästkommande)
alt. 3 Sequence snabb puls
Care kV Semi Ref.kV80 (min70kV-max120kV), CareDose4D: Quality ref mAs:500, Slide:11
Slice 1,0, Acq 128 x 0.6, Feed (34.5), Rotation time: Flex 0,28s
Scan trigger: ECG, API:Inspiration
Scan 30-80, Pulsing: Manual, Range: 35-72 Unit: %
alt. 4 Spiral
Care kV Semi Ref.kV 80 CareDose4D: Quality ref mAs:480, Slide:11
Slice 1,0, Acq 128 x 0.6, Rotation time:(0,28s), Scan trigger: ECG, API:Inspiration
Pulsing: Auto, Unit:%, EST HR.: auto